Temperature setting range |
Room temperature~100ºC |
Fluctuation range (no load) |
≤±1ºC |
Temperature distribution uniformity (no load) |
Full load ≤ ± 3 ºC, no load ≤ ± 2 ºC |
Temperature deviation (no load) |
≤±2ºC |
Temperature deviation (full load) |
≤±5ºC |
Supply voltage range |
3V ~ 3.6V adjustable |
Voltage output fluctuation |
≤ ± 0.2V |
Channel drive capacity |
≤ 4000mA, and the need to provide overcurrent protection function |
Current detection accuracy |
≤ ± 0.1mA |
Voltage detection accuracy |
≤±4mV |
With anti ESD / EOS capability |
With anti-ESD / EOS capability |
Monitoring capability |
Can real-time monitoring of aging module products SN, DDM_TX, DDM_RX, Icc, Ibias, temperature, voltage and other parameters |
Equipment functions |
Experimental temperature, current, voltage, time, over-temperature protection conditions can be freely set, data requirements can be exported in real time, the equipment aging normal operation of the box door automatically lock does not allow the opening of the box (automatic lock door way can be communicated) |
Communication interface |
With data networking upload interface, must be with LAN interface, RJ45/RS232/GPIB contains at least one of them, can be docked to our server to achieve data upload, the specific implementation mode and format can be communicated. |
Box volume |
≥ 520L |
Optical module power supply |
3.0 ~ 3.6V software adjustable, output fluctuations ≤ 1%. |
Module power supply |
DC power supply, power ≥ 2800W |
Temperature rise time |
From 20 ºC to 100 ºC <60min (empty box) |
Noise |
≤ 85dB (A) / 1m distance |
Circulating wind |
Horizontal air supply, box circulation wind > 2 m / s, when the over-temperature alarm, circulating wind shall not stop. |